METHOD
Multiparametric detection of bacterial contamination based on the photonic crystal surface mode detection
1 Laboratory of Nano-Bioengineering, National Research Nuclear University MEPhI (Moscow Engineering Physics Institute), Moscow
2 Laboratory of Spectroscopy of Condensed Matter, Institute for Spectroscopy, Russian Academy of Sciences, Troitsk
Correspondence should be addressed: Igor R. Nabiev
Kashirskoe shosse 31, Moscow, 115529; moc.liamg@veiban.rogi
Funding: this work was part of the Federal Targeted Program The National system of Chemical and Biological Security of the Russian Federation (2015-2020) supported by the Ministry of Healthcare of the Russian Federation (State grant No. K-27-НИР/144-5 dated December 24, 2015).
Acknowledgement: the authors wish to thank Tkachuk AP, Head of the Department of Translational Biomedicine (Gamaleya Research Institute of Epidemiology and Microbiology) for rabbit antibodies against the heat-labile toxin LT.
Detection of pathogenic bacteria and their toxins in liquid samples is a critical component of food and water quality control and the environmental monitoring in general. Polymerase chain reaction (PCR), a well-established technique for pathogen detection, is highly sensitive yet lengthy, reagent-consuming and labor-intensive. Besides, it cannot recognize bacterial metabolites, including protein toxins (primary pathogenicity determinants), and other specific proteins, such as prions, since it is aimed at detecting nucleic acids.At present, detection of proteins (e.g., protein toxins) in liquid samples is done by classic analytical techniques. Conventional western blotting and enzyme-linked immunosorbent assays (ELISA) are relatively user-friendly and cost-effective but at the same time low-throughput and time-consuming. They are suitable for clinical diagnosis but require additional sample preparation involving the use of enzymatic and fluorescent labels to amplify the emitted signal, which increases costs and complicates the procedure.
Fortunately, novel techniques are emerging, including electrochemical immunoassays [1–2], chemiluminescence imaging [3], fluorescence flow immunoassays with activated silica beads [4], and electrochemical assays based on the use of affibodies [5]. Advantageously, all of them are multiplex, i.e. able to detect multiple analytes in one sample. Unlike traditional approaches, these techniques are high-throughput; however, they are expensive and the yielded results are difficult to interpret, which makes them unsuitable for routine use. The majority of these techniques rely on the use of additional labels, except for the label-free electrochemical assay, a sensitive and elegant approach, which, nevertheless, cannot be multiplexed [6].
Great promise is held by surface plasmon resonance [7] based on the excitation of surface plasmon polaritons at the metal/dielectric interface by incident light. This fast and sensitive label-free technique can go without complicated sample preparation. It exploits antibodies specific for a target protein that are conjugated to the surface of a thin gold film; soluble antigens present in the liquid sample bind to the antibodies, and the resulting mass transfer is manifested as a change in the refractive index value of the superficial layer of the liquid close to the surface of the golden film.
I. Design of the experiment
We propose a novel approach to the detection of bacterial contamination based on the use of photonic crystal surface modes (PCSMs). A photonic crystal is a structure in which the refractive index (RI) is periodically modulated on the scale of a light wavelength. Such structures maintain the far-reaching longitudinal propagation of optical waves across the external surface. PCSMs are exploited in optical sensors [8–10]. PCSM-based sensors are reported to be more sensitive than those relying on surface plasmon resonance [9]. PCSMs allow detection of both parallel and perpendicular polarization of the light wave, meaning that the thickness of the surface layer and the refractive index of the liquid phase can be measured separately and a change in the reflected signal is indicative of a mass transfer event (interaction of a soluble analyte with a functionalized crystal surface) regardless of the refractive index shifts occurring in close proximity to this surface.Advantageously, this approach allows easy recovery of a chip surface by plasma cleaning. The chip can be reactivated and conjugated to another detector protein. This means that the PC chip can be used an infinite number of times and activated in advance before the actual experiment.In this work we studied the interaction between the polymer microbeads coated with a layer of a polyelectrolyte (negatively charged polyacrylic acid) and the surface of a PCSM-based sensor coated with positively charged poly(sodium 4-styrenesulfonate). Because of their size (4.08 μm in diameter), microbeads can represent a bacterial cell model. Given that, our work demonstrates the feasibility of bacterial cell detection in a liquid sample using a PCSM-based biosensor.Besides we show that a PCSM-based biosensor can detect bacterial toxins in a liquid sample using the exotoxin A of Pseudomonas aeruginosa and the heat-labile toxin LT of Escherichia coli. Below we describe the main stages in the development of the proposed analytical technique for the multiplex detection of biomarkers and its application for the detection of bacterial toxins.
II. Preparation
- Reagents
The following reagents were used: sodium chloride, > 99% (Sigma-Aldrich, S6191; USA); ethanol, 99.5% (Acros Organics, AC615090010; Belgium); acetone for high-performance liquid chromatography, 99.8% (Acros Organics, 268310010; Belgium); (3-aminopropyl)triethoxysilane (APTES), 99% (Sigma- Aldrich, 440140; USA); glutaraldehyde, grade I, 25% in H2O (Sigma-Aldrich, G5882; USA); sodium phosphate dibasic heptahydrate, 98–102% (Sigma-Aldrich, S9390; USA); sodium phosphate monobasic monohydrate, ≥ 98% (Sigma-Aldrich, S9638; USA); phosphate buffered saline (PBS), tablets (Sigma- Aldrich, P4417; USA); protein A, United States Pharmacopeia (USP) Reference Standard (Sigma-Aldrich, 1578805; USA); poly(sodium 4-styrenesulfonate) (PSS), average Mw ~70,000 (Sigma-Aldrich, 243051; USA); poly(allylamine hydrochloride) (PAH), average Mw ~50,000 (Sigma-Aldrich, 283223; USA); poly(acrylic acid) (PAA), average Mw ~130,000 (Sigma- Aldrich, 181293; USA); rabbit antibodies to the exotoxin A of Pseudomonas aeruginosa, whole antiserum (Sigma-Aldrich, P2318; USA); the exotoxin A of Pseudomonas aeruginosa, lyophilized powder (Sigma-Aldrich, P0184; USA); the heat-labile recombinant toxin LT of Escherichia coli, subunit B, > 90%, lyophilized powder (Sigma-Aldrich, E8656; USA); crystalline bovine serum albumin (BSA), lyophilized powder, ≥ 98.0% (GE) (Sigma-Aldrich, 05470; USA).The rabbit antibodies against the heat-labile toxin LT were courtesy of Tkachuk AP, Head of the Department of Translational Biomedicine, Gamaleya Research Institute of Epidemiology and Microbiology.
- A biosensor based on the detection of surface modes in a photonic crystal
We used the PCSM-based biosensor EVA 2.0 described in [10] with the following one-dimensional PC structure: substrate/ (LH)3L'/water, where L is a SiO2 layer with thickness d1 = 186.4 nm, H is a Ta2O5 layer with thickness d2 = 115.2 nm, and L' is a SiO2 layer with thickness d3 = 776.8 nm. The 7-layer structure with the starting and finishing layers made of SiO2 was created by magnetron sputtering. The prism and the glass plate were fabricated from BK-7 glass. Refractive indices of the substrate, SiO2, Ta2O5 and water at λ = 632.8 nm were n0 = 1.515, n1 = n3 = 1.47, n2 = 2.07, and ne = 1.332, respectively.The obtained data were processed in Origin 8.1 (OriginLab; USA).
- Microbeads
Latex microbeads of 4.08 μm in diameter (MF-COOH-AR421 Carboxyl-modified Melamine Resin-Research Particles, 10 ml 10% w/v aq. suspension (microParticles GmbH; Germany)) were coated with alternating layers of the oppositely charged polyelectrolytes PAH and PSS using the layer-by-layer deposition technique proposed in [11] following the pattern PAH/PSS/ PAH/PSS/PAH; the coating was finished with a layer of PAA.
- Preparation of the photonic crystal silica chip for the experiment with microbeads
Before the experiment, the silica PC chip was left in the UV cleaner box for at least 30 min. The chip was mounted in the microfluidic flow cell for further measurements. To make the liquid flow through the cell, a peristaltic pump was used. After the flow cell was assembled, 1% aqueous APTES solution was pumped through it at 100 μl/min for 5 min to activate the surface of the PC by amino groups. Then 100 mg/ml solutions of PSS and PAH in 0.5 NaCl were pumped through the flow cell successively for 5 min each. Before changing the solution, the flow cell was washed with 0.5 NaCl. To ensure even coating of the chip, three bilayers of PSS/PHH were applied onto it in succession. After applying the finishing layer of PAH, the medium was replaced with 100 mM sodium phosphate buffer (pH 8.0) containing 0.5 M NaCl. The microbeads were suspended in the same buffer at a concentration of 106 beads per 1 ml. The microbead suspension was pumped through the flow cell at 100 μl/min for 10 min.
- Preparation of the photonic crystal silica chip for the experiments with proteins
The PC chip was left in the UV cleaner box for 30 min to clean each side of the crystal and then washed once in acetone and once in ethanol successively followed by 3 washes in milliQ, with simultaneous exposure to UV light for 5 min. After that, the substrate was left to dry at 70 °C. The cleaned chip was incubated in 1% APTES solution in acetone for 16 h to functionalize its surface by amino groups and then calcinated at 120 °С for 90 min. Such aminized chips can be stored for at least one month without losing their properties. Before taking measurements, the chip was treated with 2.5% glutaraldehyde in the sodium-phosphate buffer (pH 7.4) under gentle stirring for 30 min. Dextran or other polymer frameworks were not used for chip preparation. The flow cell was attached to the prepared chip. The flow speed was 30 μl/min. To functionalize the surface of the chip with detector antibodies, the solution of protein A (50 μg/ml) in PBS was pumped through the cell until the signal stabilized. Then the flow cell was washed in PBS for 2 min. After the monolayer of protein A bound to the surface of the chip, the antibody solution (50 μg/ml) was pumped through the flow cell again until the signal was stable. Protein A present on the surface of the PC chip bound the Fc fragment of the immunoglobulin ensuring a uniform orientation pattern of bound antibodies. The surface regions that did not react with the protein were neutralized with 2% solution of bovine serum albumin (BSA) in PBS, and the flow cell was washed with PBS until the signal reached the plateau. Finally, the studied sample was pumped though the cell at 30 μl/min.
III. Deposition of polyelectrolyte-coated microbeads on the chip coated with a layer of an oppositely charged polyelectrolyte
Layer-by-layer deposition of polyelectrolytes on the APTES-activated chip surface caused a stepwise increase in the thickness of the superficial layer (fig. 1A). Thickness values observed for the first PSS and PAH layers were lower than those observed for the subsequent layers, suggesting that two first layers of polyelectrolytes covered the surface incompletely. The binding of PAA-coated microbeads to the surface of the crystal caused the superficial layer to slowly grow in thickness. This phenomenon was not observed if the buffer solution pumped through the flow cell did not contain microbeads.Notably, the kinetics of microbead binding was linear (R2 = 0.996) and not dose dependent (fig. 1B), possibly due to a low concentration of microbeads unable to saturate the chip surface (the area of the chip surface in contact with the solution was 1 cm2, as determined by the flow cell architecture). The area on the chip surface occupied by one microbead did not exceed 13 μm2, i.e. to fully cover the surface one would need ~107 microbeads. Therefore, the thickness of the microbead monolayer could be expected to be equal to the diameter of the microbead, i.e. 4.08 μm. Accordingly, the binding of 106 microbeads would correspond to a 400 nm thickness of the superficial layer. However, at low thickness values we can only measure the optical depth of a layer, that is its physical thickness multiplied by the refractive index. By taking the refractive index equal to 1.56 [10], we get the expected thickness of 624 nm consistent with the observed thickness. Knowing the flow rate, the rate of superficial thickness growth and the concentration of microbeads, we can estimate the number of the bound microbeads. As the microbeads are pumped through the cell for 10 min, the superficial layer increases in thickness by 0.57 nm, meaning that 1 × 103 microbeads (0.1% of the total number of pumped microbeads) have been bound.
IV. Detection of bacterial toxins
To attempt detection of protein toxins in a solution using the PCSW-based biosensor, we selected two toxins: the exotoxin of Pseudomonas aeruginosa and the heat-labile toxin of Escherichia coli. Pseudomonas aeruginosa is a gram-negative motile rod-shaped bacterium sized 1–5 μm by 0.5–1 μm. It inhabits water and soil and causes opportunistic and nosocomial infections in humans. The exotoxin A of the bacteria consists of three domains: binding, transport and catalytic. Its cytotoxic activity is determined by the catalytic domain that inhibits synthesis of proteins by ATP-ribosylation of the elongation factor EF-2 [12–13]. Detection of this exotoxin is an important diagnostic problem normally solved by PCR which identifies the gene coding for it [14], ELISA which detects the presence of this toxin and a few other analytical techniques. Escherichia coli is a gram-negative rod-shaped bacterium colonizing the lower intestinal tract of warm-blooded organisms. Enterotoxigenic strains of Escherichia coli can cause diarrhea by producing toxins such as the heat labile LT toxin. The latter consists of two subunits: the catalytically active subunit A and the subunit B responsible for the binding to the cell [15]. LT expression stimulates secretion and facilitates the binding of the bacteria to epithelial cells [16].For toxin detection, we developed a system consisting of flow cells with 1, 2 or 4 channels. Briefly, the glass chamber with 1, 2 or 4 grooves is connected to a chemically activated sensitive surface of the silica chip functionalized with detector molecules. The amplitude of the signal recorded during antigen-antibody interactions on the sensor surface is directly proportional to the number of binding sites on the surface. The surface regions that have not reacted with the antigen are coated with BSA to avoid emission of a nonspecific signal. If a cell has 2 or 4 channels, 1 of them can be used as a negative control; its surface is coated with BSA instead of antibodies.
In the experiment, the 50 μg/ml solution of the exotoxin A of Pseudomonas aeruginosa in PBS was pumped through the flow cell at 100 μl/min (fig. 2). The obtained curve was similar to the SPR sensorgram and followed the same pattern of kinetics [17]. The binding of the protein to the sensor surface caused a rapid increase in the thickness of the superficial layer, which reaches the plateau quite soon. The plateau represents the equilibrium between the absorption and desorption rates. The kinetic curve can be approximated using the Langmuir absorption model. It has been theorized that the detection limit allowed by PCSM is 70 fg/mm2 of the studied sensor surface, which is well above the sensitivity of surface plasmon resonance [18]. But in practice, the lowest detectable concentration of exotoxin A in our experiment was 30 μg/ml, in spite of the low levels of the background noise. The sensorgram recorded during the detection of the heat-labile toxin LT (50 μg/ml in PBS) demonstrated that the detection limit was 10 μg/ml (fig. 3). Of note, in the experiment we used whole antiserum. The use of affinity purified antibodies will definitely improve the sensitivity of the biosensor allowing it to reach the anticipated value [18], which is way higher than the sensitivity of surface plasmon resonance.
V. Kinetics of bacterial cell binding to the surface of the sensor functionalized with antibodies
The upper limit of the binding rate for the microbeads was calculated using the Smoluchowski-Levich equation [19] and was 3.46 particles/cm2; the working surface of the sensor was 1 cm2, so the calculated value meant that 2,000 particles would be bound within 10 min. Here we can conclude that the binding rate of polyelectrolyte-coated microbeads to the surface of the biosensor coated with the oppositely charged polyelectrolyte in our system was close to the highest possible rate. Perhaps, it did not reach its maximum because the binding of microbeads to the surface is not instant and under real life conditions is determined not only by the diffusion of microparticles but also by their interaction with the surface. Therefore, we assume that the kinetics of bacterial cell binding to the sensor surface functionalized by antibodies can differ from that demonstrated in our experiment involving microbeads. It should be noted that changes in the composition of the microbeads-containing solution affected the binding kinetics only insignificantly.
VI. Optimization of detection of bacterial toxins
The results obtained in the experiment demonstrate the feasibility of multiplex detection of bacterial toxins using a PCSM-based biosensor. The detection limit can be lowered by (1) the use of affinity purified antibodies to the studied toxins; the antibody-antigen affinity is crucial because each molecule lacking it negatively affects the effective area of the sensor surface; (2) changing the parameters of the measuring flow cell [20]; (3) enhancing the signal by gold nanoparticles: the intensity of the signal from the crystal depends on mass transfer, so the use of specific antibodies recognizing a sterically distant epitope of the target protein and conjugated to colloid gold will increase the signal intensity at the same concentration of an analyte [21–22].
CONCLUSIONS
This work demonstrates the feasibility of PCSM-based detection of microparticles similar to bacterial cells in terms of their size in a liquid sample. We hope that the obtained results will assist development of protocols for real-time detection of bacteria in flow cells. The proposed technique is suitable for multiplex detection of protein toxins: the exotoxin A of Pseudomonas aeruginosa and the heat-labile toxin LT of Escherichia coli. Our findings could be used to design rapid and cheap tools for measuring bacterial contamination of water and food and establishing fast and accurate diagnosis of bacterial infections by detecting bacterial cells and their metabolites in the samples of biological fluids.